Fluke-1507, 1503 절연저항 미터

절연저항 테스터기

원격프루브를 사용하여 프루브에서 절연테스트를 할 수 있습니다.


*절연저항 측정

1507: 0.01 MΩ to 10 GΩ 
1503: 0.1 MΩ to 2000 MΩ

* 테스트 전압

1507: 50 V, 100 V, 250 V, 500 V, 1000 V 
1503: 500 V, 1000 V

* 전압측정 : 0.1 V to 600 V

* 전류측정 : 200 mA

* 저항측정 : 0.01 Ω to 20.00 kΩ 

* 오토OFF 기능

* 30V 이상에서 측정시 경고

 

FLUKE 1503/1507 절연테스터는 1000V(가변) 테스트 레인지를 
가진 절연저항계 입니다. 전기설비, 전동기, 발전기, 케이블 등의
절연 테스트, 수리, 점검 어플리케이션에 적합한 테스트 장비 입니다. 

 

AC/DC Voltage Measurement

Accuracy

Range

600.0 V

Resolution

0.1 V

50 Hz to 400 Hz ± (% of Rdg + Digits)

± (2 % + 3)

Input impedance

3 MΩ (nominal), < 100 pF

Common mode rejection ratio (1 kΩ unbalanced):

> 60 dB at dc, 50 or 60 Hz

Overload Protection

600 V rms or dc

 

Earth bond resistance measurement

Range/Resolution

20.00 Ω

0.01 Ω

200.0 Ω

0.1 Ω

2000 Ω

1.0 Ω

20.00 kΩ

0.01 kΩ

Accuracy

± (1.5 % + 3)

Overload Protection

2 V rms or dc

Open Circuit Test Voltage

> 4.0 V, < 8 V

Short circuit current

> 200.0 mA

 

 

Earth bond resistance measurement

Range/Resolution

20.00 Ω

0.01 Ω

200.0 Ω

0.1 Ω

2000 Ω

1.0 Ω

20.00 kΩ

0.01 kΩ

Accuracy

 

± (1.5 % + 3)

Overload Protection

 

2 V rms or dc

Open Circuit Test Voltage

 

> 4.0 V, < 8 V

Short circuit current

 

> 200.0 mA

 

General Specifications

Maximum voltage applied to any terminal:

600 V ac rms or dc

Storage Temperature

-40 °C to 60 °C (-40 °F to 140 °F)

Operating Temperature

-20 °C to 55 °C (-4 °F to 131 °F)

Temperature Coefficient

0.05 x (specified accuracy) per °C for temperatures < 18 °C or > 28 °C (< 64 °F or > 82 °F)

Relative Humidity

0 % to 95 % @ 10 °C to 30 °C (50 °F to 86 °F)

0 % to 75 % @ 30 °C to 40 °C (86 °F to 104 °F)

0 % to 40 % @ 40 °C to 55 °C (104 °F to 131 °F)

Vibration

Random, 2 g, 5-500 Hz per MIL-PRF-28800F, Class 2 instrument

Shock

1 meter drop per IEC 61010-1 2nd Edition (1 meter drop test, six sides, oak floor)

Electromagnetic compatibility:

In an RF field of 3 V/M, accuracy = specified accuracy (EN 61326-1:1997)

Safety

Complies with ANSI/ISA 82.02.01 (61010-1) 2004, CAN/CSA-C22.2 NO. 61010-1-04, and IEC/EN 61010-1 2nd Edition for measurement category IV 600 V (CAT IV)

Certifications:

CSA per standard CSA/CAN C22.2 No. 61010.1-04; TUV per standard IEC/EN 61010-1 2nd Edition

Batteries

Four AA batteries (NEDA 15A or IEC LR6)

Battery Life

Insulation test use:

Tester can perform at least 1000 insulation tests with fresh alkaline batteries at room temperature. These are standard tests of 1000 V into 1 MΩ with a duty cycle of 5 seconds on and 25 seconds off.

Resistance measurements:

Tester can perform at least 2500 earth bond resistance measurements with fresh alkaline batteries at room temperature. These are standard tests of 1 Ω with a duty cycle of 5 seconds on and 25 seconds off.

Size

5.0 cm H x 10.0 cm W x 20.3 cm L (1.97 in H x 3.94 in W x 8.00 in L)

Weight

550 g (1.2 lb)

IP Rating

IP40

Altitude

Operating:

2000 m CAT IV 600 V, 3000 m CAT III 600 V

Non-operating (storage):

12,000 m

Over-range capability:

110 % of range

Included accessories:

TL224 Test Leads, TP74 Test Probes, clips PN 1958654 (red) and PN 1958646 (black), holster and remote probe